Acquisition

 

- Choose best septal view for LAO then RAO with - 45° and also a lateral with + 45°, in respect to the septal view.

- Obtain 16-32 gated frames per cardiac cycle acquisition.

- Collect each view for 3-7 million counts at resting condition with high resolution collimator.

- In case of exercise obtain 3 min acquisition according to Bruce or ECG protocol, plus baseline and postexercise views.

- Spect may be an option but it is  not widely used.

References:

1 Flotats A et al. Cardiologia Nuclear en la Insuficiencia Cardíaca, in Medicina Nuclear Aplicaciones Clínicas. Ignasi Carrio y Patricio González . Editorial Masson, Barcelona España, 2003, p 87-104.

2 Wittry Mark D. Et al, Procedure Guideline for Gated Equilibrium Radionuclide Ventriculography 2.0. Ed by Society of Nuclear Medicine, 2001-2002, p 1-6.

3 Daou D. SPECT radionuclide angiography: it is time for a consensus statement. Eur J Nucl Med Mol Imaging. 2007 Nov;34(11):1729-34.
 